6 53220A/53230A Math, Graphing, and Data Logging
228 Keysight 53220A/53230A User’s Guide
value - directly specifies the histogram lower and upper range values. The value
ranges are:
-1.0E+15 to -1.0E-15, 0.0, 1.0E-15 to 1.0E+15
The default lower and upper ranges and the ranges set following a reset (*RST) or
instrument preset (SYSTem:PRESet) are 0.0.
When setting the ranges, both a lower range and an upper range must be
specified. Any of the following steps will prevent a “settings conflict” error (e.g.
lower range greater than upper range) due to the order in which ranges are set:
setting the upper range first
setting both ranges on the same line programmatically
–enabling the histogram after the ranges are set
setting the ranges automatically
Setting the bin ranges to min and max
With AutoRange Off, the Set to Min/Max softkey is another method for setting the
histogram bin ranges (lower and upper). Depending on the instrument state, Set to
Min/Max will set the ranges as follows:
When ‘Statistics’ (under the Math key) are enabled, the minimum and
maximum values from the reading statistics are used.
If readings are in progress and ‘Statistics’ are off, the instrument will select a
minimum and a maximum value from (up to) the last 10,000 readings taken.
If no previous readings are available, a minimum bin range of 0 and a
maximum bin range of 1 000 000 000.0 are set.