131
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide
7 Digital Channels
To connect the digital probes to the device under test / 131
Acquiring waveforms using the digital channels / 135
To display digital channels using Autoscale / 135
Interpreting the digital waveform display / 136
To switch all digital channels on or off / 138
To switch groups of channels on or off / 138
To switch a single channel on or off / 138
To change the displayed size of the digital channels / 137
To reposition a digital channel / 139
To change the logic threshold for digital channels / 138
To display digital channels as a bus / 140
Digital channel signal fidelity: Probe impedance and grounding / 143
This chapter describes how to use the digital channels of a Mixed-Signal
Oscilloscope (MSO).
The digital channels are enabled on MSOX3000T X-Series models and
DSOX3000T X-Series models that have the MSO upgrade license installed.
To connect the digital probes to the device under test
1 If necessary, turn off the power supply to the device under test.
Turning off power to the device under test would only prevent damage that
might occur if you accidentally short two lines together while connecting
probes. You can leave the oscilloscope powered on because no voltage appears
at the probes.