EasyManuals Logo

Keysight DSOX3034T User Manual

Keysight DSOX3034T
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #258 background imageLoading...
Page #258 background image
258 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
15 Measurements
• “Delay" on page 261
• “Phase" on page 262
• “X at Min Y" on page 264
• “X at Max Y" on page 264
Period
Period is the time period of the complete waveform cycle. The time is measured
between the middle threshold points of two consecutive, like-polarity edges. A
middle threshold crossing must also travel through the lower and upper threshold
levels which eliminates runt pulses. The X cursors show what portion of the
waveform is being measured. The Y cursor shows the middle threshold point.
Frequency
Frequency is defined as 1/Period. Period is defined as the time between the
middle threshold crossings of two consecutive, like-polarity edges. A middle
threshold crossing must also travel through the lower and upper threshold levels
which eliminates runt pulses. The X cursors show what portion of the waveform is
being measured. The Y cursor shows the middle threshold point.
See Also • “To isolate an event for frequency measurement" on page 258
To isolate an event for frequency measurement
The following figure shows how to use Zoom mode to isolate an event for a
frequency measurement.
You may need to change the measurement window setting so that the
measurement is made in the lower, Zoom window. See “Measurement
Window" on page 272.
If the waveform is clipped, it may not be possible to make the measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DSOX3034T and is the answer not in the manual?

Keysight DSOX3034T Specifications

General IconGeneral
BrandKeysight
ModelDSOX3034T
CategoryTest Equipment
LanguageEnglish

Related product manuals