32 SENT Triggering and Serial
Decode
Setup for SENT Signals / 497
SENT Triggering / 502
SENT Serial Decode / 504
SENT (Single Edge Nibble Transmission) triggering and serial decode requires the
DSOXT3SENSOR license.
Setup for SENT Signals
To set up the oscilloscope to capture SENT signals:
1 Connect an oscilloscope channel to the signal in the device under test.
Analog channels or digital channels can be used.
2 Press [Serial].
3 Press the Serial softkey, turn the Entry knob to select the desired slot (Serial 1 or
Serial 2), and press the softkey again to enable decode.
4 Press the Mode softkey; then, select SENT.
5 Press the Source softkey to open the SENT Source Menu.