13 Acquisition Control
If the signal you are measuring has more than about 1 s of inactivity, consider
selecting the Normal trigger mode to prevent AutoTriggering. See “To select the
Auto or Normal trigger mode" on page 208.
See Also • “Navigating Segments" on page 230
• “Measurements, Statistics, and Infinite Persistence with Segmented
Memory" on page 231
• “Segmented Memory Re-Arm Time" on page 231
• “Saving Data from Segmented Memory" on page 231
Navigating Segments
1 Press the Current Seg softkey and turn the Entry knob to display the desired
segment along with a time tag indicating the time from the first trigger event.
You can also navigate segments using the [Navigate] key and controls. See “To
navigate segments" on page 76.