EasyManuals Logo

Keysight DSOX3034T User Manual

Keysight DSOX3034T
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #295 background imageLoading...
Page #295 background image
295
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide
18 Frequency Response
Analysis
To make connections / 295
To set up and run the analysis / 296
To view and save the analysis results / 298
Frequency Response Analysis (FRA) is a licensed feature you can get by
purchasing the DSOXT3FRA upgrade.
The Frequency Response Analysis (FRA) feature controls the built-in waveform
generator to sweep a sine wave across a range of frequencies while measuring the
input to and output from a device under test (DUT). At each frequency, gain (A)
and phase are measured and plotted on a frequency response Bode chart.
When the frequency response analysis completes, you can move a marker across
the chart to see the measured gain and phase values at each frequency point. You
can also adjust the chart's scale and offset settings for the gain and phase plots.
To make connections
The waveform generator output is connected to a device under test (DUT). The
input to the device and the output from the device are probed by the
oscilloscope's input channels.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DSOX3034T and is the answer not in the manual?

Keysight DSOX3034T Specifications

General IconGeneral
BrandKeysight
ModelDSOX3034T
CategoryTest Equipment
LanguageEnglish

Related product manuals