EasyManuals Logo

Keysight DSOX3034T User Manual

Keysight DSOX3034T
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
194 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
11 Triggers
The Generic selection lets you trigger on custom bi-level and tri-level sync video
standards. See “To set up Generic video triggers" on page 195.
7 Press the Auto Setup softkey to automatically set up the oscilloscope for the
selected Source and Standard:
• Source channel vertical scaling is set to 140 mV/div.
• Source channel offset is set to 245 mV.
• Source channel is turned on.
• Trigger type is set to Video.
• Video trigger mode is set to All Lines (but left unchanged if Standard is
Generic).
• Display Grid type is set to IRE (when Standard is NTSC) or mV (see “To select
the grid type" on page 158).
• Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM standards
or 4 µs/div for EDTV or HDTV standards (unchanged for Generic).
• Horizontal delay is set so that trigger is at first horizontal division from the
left (unchanged for Generic).
You can also press [Analyze]> Features and then select Video to quickly access the
video triggering automatic set up and display options.
8 Press the Mode softkey to select the portion of the video signal that you would
like to trigger on.
The Video trigger modes available are:
• Field1 and Field2 — Trigger on the rising edge of the first serration pulse of
field 1 or field 2 (interlaced standards only).
• All Fields — Trigger on the rising edge of the first pulse in the vertical sync
interval.
• All Lines — Trigger on all horizontal sync pulses.
• Line — Trigger on the selected line number (EDTV and HDTV standards only).
• Line: Field1 and Line:Field2 — Trigger on the selected line # in field 1 or field 2
(interlaced standards only).
• Line: Alternate — Alternately trigger on the selected line # in field 1 and field 2
(NTSC, PAL, PAL-M, and SECAM only).
9 If you select a line # mode, press the
Line # softkey, then rotate the Entry knob
to select the line number on which you want to trigger.