236 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
14 Cursors
When individual cursors are selected, press the softkey to disable or reenable
the cursor display.
6 To change the cursor units, press the Units softkey.
In the Cursor Units Menu:
You can press the XUnits softkey to select:
• Seconds (s).
• Hz (1/s).
• Phase (°) — when selected, use the Use X Cursors softkey to set the current X1
location as 0 degrees and the current X2 location as 360 degrees.
• Ratio (%) — when selected, use the Use X Cursors softkey to set the current X1
location as 0% and the current X2 location as 100%.
You can press the YUnits softkey to select:
• Base — the same units used for the source waveform.
• Ratio (%) — when selected, use the Use YCursors softkey to set the current Y1
location as 0% and the current Y2 location as 100%.
For phase or ratio units, once the 0 and 360 degree or 0 and 100% locations are
set, adjusting cursors will display measurements relative to the set locations.
7 Adjust the selected cursor(s) by turning the Cursors knob.
You can also position cursors using the touchscreen. See “Drag Cursors" on
page 51.