400 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
25 CAN/LIN Triggering and Serial Decode
3 Press the Trigger on: softkey; then, turn the Entry knob to select the trigger
condition:
• SOF - Start of Frame — triggers at the start bit for both data and overload
frames.
• EOF - End of Frame — triggers at the end of any frame. *
• Frame ID — triggers on any standard CAN (data or remote) or CAN FD frame
at the end of the 11- or 29-bit ID field.
• Data Frame ID (non-FD) — triggers on standard CAN data frames at the end of
the 11- or 29-bit ID field.
• Data Frame ID and Data (non-FD) — triggers on any standard CAN data frame at
the end of the last data byte defined in the trigger. The DLC of the packet
must must match the number of bytes specified.
• Data Frame ID and Data (FD) — triggers on CAN FD frames at the end of the last
data byte defined in the trigger. You can trigger on up to 8 bytes of data
anywhere within the CAN FD data, which can be up to 64 bytes long.
• Remote Frame ID — triggers on standard CAN remote frames at the end of the
11- or 29-bit ID field.
• Error Frame — triggers after 6 consecutive 0s while in a data frame, at the
EOF. *
• Acknowledge Error — triggers on the acknowledge bit if the polarity is
incorrect. *
• Form Error — triggers on reserved bit errors. *
• Stuff Error — triggers on 6 consecutive 1s or 6 consecutive 0s, while in a
non-error or non overload frame. *
• CRC Field Error — triggers when the calculated CRC does not match the
transmitted CRC. In addition, for FD frames, will also trigger if the Stuff
Count is in error. *
• Spec Error (Ack or Form or Stuff or CRC) — triggers on Ack, Form, Stuff, or CRC
errors. *
• All Errors — triggers on all Spec errors and error frames. *
• BRS Bit (FD) — triggers on the BRS bit of CAN FD frames. *
• CRC Delimiter Bit (FD) — triggers on the CRC delimiter bit in CAN FD frames. *