Keysight E5092A ENA Series Configurable Multiport Test Set 211
Post-Repair Procedures
Post-Repair Procedures
Switch 3 (20GHz
SP4T Pin Switch
Module)
(P9398-60001)
No adjustment
needed
Perform the following performance tests.
- Write SN
- Load Match, Insertion Loss and
Isolation Test for Switch 3 (SP4T, Port3,
3A, 3B, 3C and 3D)
- LED Check
- DCBUS Check
Switch 4 (20GHz
SP4T Pin Switch
Module)
(P9398-60001)
No adjustment
needed
Perform the following performance tests.
- Write SN
- Load Match, Insertion Loss and
Isolation Test for Switch 4 (SP4T, Port4,
4A, 4B, 4C and 4D)
- LED Check
- DCBUS Check
Table 6-1 Post-Repair Procedures
Replaced
Assembly or Part
Required
Adjustments
Correction Constants
(CC)
Verification

Table of Contents