Appendix page 107
Error No. Fault Appearance Remedy
35
Underload (force
pressing against
balance from
below is too
large).
Appears if attaching and
removing the weight when
adjusting the instrument or
if the motor presses the
measuring body against the
balance.
Remove load and
press ESC key to
switch instrument off
and on.
49
Motor at lower
limit.
On switch-on
: only
information for the operator.
During measurement
: no
film tearing is recorded; the
sample carrier moves
downwards to the limit.
Briefly
press the ESC
key.
The measurement is
terminated. Confirm
with ESC and restart
the measurement
50
Motor at upper
limit.
On switch-on
: only
information for the operator.
During measurement no
film tearing is recorded or
no surface is found. The
sample carrier moves
upwards to the limit.
Briefly
press the ESC
key.
The measurement is
terminated. Confirm
with the ESC key and
start the measure-
ment again. If
necessary, ensure
that the surface is
higher.
51
Motor cannot be
initialized.
A fault exists. Contact our technical
support.
52
Motor transmits
no data.
A fault exists. Contact our technical
support.
53
Critical internal
motor fault
(overload)
Motor is overloaded; e.g.
when an object gets
jammed.
Remove the cause of
the overload.
65
Lamella torn Lamella tears when
maximum force is being
determined.
Briefly
press the ESC
key and restart the
measurement. If the
message appears
again reduce the
measuring speed.