page 106 Appendix
Light density: not relevant with linear correction (4)
Relaxation: not relevant with linear correction (4)
Min. standard deviation: not relevant because
Reading limit
= 1
Number of values for SD: not relevant because
Reading limit
= 1
III Error messages
If an error occurs in controlling the instrument or when carrying out a measurement
then a corresponding error message will appear on the display. The possible error
messages and the steps to be taken to remedy them are given in the following list.
Error No. Fault Appearance Remedy
17
Clock not
working.
Occurs before the time has
been set for the first time
and after the battery has
been exchanged.
Set the current time
and date (
Ö ch. 10.4)
18
Battery was
empty and has
been exchanged,
standard values
set.
Only occurs immediately
after the battery has been
changed.
Only for information;
requires no action.
Does not appear the
next time the
instrument is switched
on.
19
Battery empty. According to the
manufacturer the battery
has a life of up to 10 years;
however, usually the
battery has to be
exchanged earlier.
Exchange the battery
in the K11 controller
PCB. Contact our
technical customer
service if necessary.
33
Force sensor not
working (no data
transmission).
Instrument is not steady
enough.
Find a vibration-free
location.
34
Overload (force
acting
downwards is too
large).
Appears if attaching and
removing the weight when
adjusting the instrument.
Remove load and
press ESC key.