Appendix page 105
II Settings for compatibility with the Tensiometer K10
Customers who own K10 Tensiometers can set the parameters for the K11 so that it
behaves in exactly the same way as a K10. However, this possibility is only provided to
allow comparative measurements. If you wish to measure with the greatest possible
degree of accuracy then the parameters listed here are unsuitable.
Settings for simulation of K10 measurements: SFT Plate
Measuring speed = 4.2 mm/min
Immersion speed = 8.4 mm/min.
Detection speed = 4.2 mm/min.
Gravitational acceleration = 9.80665 m/s
2
Immersion depth = 2 mm
Sensitivity = 0.008 g
Min. standard deviation: not relevant because
Reading limit
= 1
Number of values for SD: not relevant because
Reading limit
= 1
Measuring mode = 0 (SFT)
Reading limit = 1
Interval: not relevant because
Reading limit
= 1
Wetted Length = 40.200mm
Settings for simulation of K10 measurements: SFT Ring
Measuring speed = 4.2mm/min
Immersion speed = 100mm/min
Search speed = 15mm/min
Gravitational acceleration = 9.80665 m/s
2
Immersion depth = 3 mm
Sensitivity = 0.003 g
Min. standard deviation: not relevant because
Reading limit
= 1
Number of values for SD: not relevant because
Reading limit
= 1
Measuring mode = 0 (SFT)
Correction = 4 (linear)
Reading limit = 1
Heavy density: not relevant with linear correction (4)