Basic settings (setup menu and special functions) page 73
*
With the setting
auto
(number 3) the Firmware will automatically check which
temperature sensors are available and select the appropriate sensors
automatically. If both sensors are available then the external sensor will be selected
automatically.
*
The selected sensor can be seen at on the display. The temperature is shown in
the field at the top right; below the ° character an
I
stands for the internal and
E
for
the external sensor.
*
If no thermostat jacket but the platform is installed then only the external
temperature sensor is available.
*
If the selected temperature sensor has not been connected then question marks
appear instead of the temperature.
10.3.4 Language / Sprache
The language in which the parameters and information appears in the display is
entered here. You can choose between
English
(0) and
German
(1).
*
The chosen language is also used for labelling the parameters and measured
values in the printer output and in the data logger add-in of the software LabDesk.

Table of Contents

Related product manuals