EasyManuals Logo

kruss K11 MK3 User Manual

Default Icon
115 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Operating the instrument page 21
To switch the K11 to standby mode: Press the key
for at least two seconds.
4.1.3 Resetting the instrument
If the K11 no longer responds to the Control Panel: Press the
RESET
button on the
back of the K11.
4.1.4 Warming-up time
!
After the instrument has been switched on the force sensor must reach its
operating temperature before measurements can be started. This requires a
warming-up time of at least 2 hours.
*
In the standby mode the balance retains its operating temperature; this means that
no warming-up time is required when the instrument is switched on with the ESC
key. This is why the instrument should not be switched off completely during a
working day.
4.2 Error message behaviour
Ö
You will find a complete list of possible error messages in the appendix on page
106.
Switch off the K11 with the key .
Remedy the error and switch on the K11 again.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the kruss K11 MK3 and is the answer not in the manual?

kruss K11 MK3 Specifications

General IconGeneral
Brandkruss
ModelK11 MK3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals