EasyManuals Logo

LeCroy WaveAce series User Manual

Default Icon
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
WaveAce Series
35
WA-GSM-E RevB
1. Choose an input coupling mode. Turn on the desired channel by
pressing the appropriate channel button - CH1 or CH2 (or CH3 or 4
on 4 channel models).
2. Now, press the Coupling option button on page 1/2 of the channel
menu, and then select a coupling mode from the menu.
Limiting Bandwidth
You can limit the bandwidth to reduce display noise. When you turn
Bandwidth Limit ON, the Bandwidth Limit value is set to 20 MHz. It also
filters the signal to reduce noise and other unwanted high frequency
components.
Note: The oscilloscope vertical response rolls off slowly above its
bandwidth; or, above 20 MHz when the Bandwidth Limit is set to ON.
Therefore, the FFT spectrum can show valid frequency information
higher than the oscilloscope bandwidth. However, the magnitude
information near or above the bandwidth will not be accurate.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy WaveAce series and is the answer not in the manual?

LeCroy WaveAce series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8-bit
Input CouplingAC, DC, GND
Power Supply100-240 VAC, 50/60 Hz
Sample Rate2 GS/s
Display800 x 480
Input Impedance1 MΩ
Timebase Range5 ns/div to 50 s/div
Trigger TypesEdge, Pulse Width, Video, Slope
ConnectivityUSB host, USB device