Recommendations
To avoid unintended activation of actuators if hand control cables short-circuit, LINAK recommends to use an OpenBus system (CO65).
If there is a risk that the system is overloaded and therefore shuts down thermally, LINAK recommends to use quick release actuators. These will allow
functions to be lowered manually in case of a CO65 malfunction due to misuse/abuse.
If the customer has other essential performance than “no unintended movement”, he must consider this in his own risk analysis. LINAK disclaims
any liability.
If the actuator or the control cable is removed from the control box, the cable lock must be applied. To ensure movement in this case, LINAK
recommends to use quick release actuators in the application.
To avoid cables from being damaged by pulling, LINAK recommends to make safe cabling. If movement is an essential performance, LINAK
recommends to apply quick release actuators, for example, to ensure movement.
To avoid thermal protection from being activated, do not exceed load specifications. If movement is an essential performance, LINAK recommends
to apply quick release actuators, for example, to ensure movement.
Sales must request a review of the products according to current cut-off limits.
Push plugs fully into correct sockets. Make sure that the plugs are completely inserted.
Mount control box lid and close lid until locked in place.
Motor cable
Always use 6-wire cables.
Please note that angled motor cable plugs are required for connection to the control box.
Warnings
Use EPR or ensure that the user takes care not to squeeze the mains cable.
Always check correct assembly after mounting and service to ensure that the cable lock is mounted. (Connectors are usually removed during cleaning)
Always use approved chemicals with the housing as the plastic may show corrosion caused by some chemicals. As a result water may accumulate/
gather in housing.
Take special precautions concerning 3rd party interfacing. Please contact LINAK for further information.
Make a review of all product specifications before system set-up if the current cut-off limit is higher than the maximum allowed current cut-off for
the actuator.
To avoid cable interruption and actuator defects make a proper cable installation and inspect regularly for wear and damage. Defective parts must
be replaced.
After service inspection, the application must be tested for correct functionality before it is put into operation to avoid actuator plugs being mixed
during service. Operators must not be inside entrapment area.
To avoid electrical failure or system disturbance inspect regularly for wear and damage. Defective parts must be replaced.
Make a proper cable installation to avoid short-circuit cables for handset/controls. Regular inspection must be made for wear and damage. Defective
parts must be replaced.
Do not mount the actuator with the spindle facing downwards to avoid that the actuator slips off the bracket with mounted control box. The bracket
can come loose when exposed to shock or hard vibratio, for instance when passing doorsteps. Regular inspection must be made to ensure proper
fixation of control box and bracket on actuator.