The CHL40 charger is an important part of the LIFT40 family. It functions
as a wall charger when mounted on the wall, but also as a more mobile
charger simply placed horizontally on its back on any surface.
3. CHL40 (MEDLINE
®
CARELINE
®
)
Usage
Usage temperature: +5 °C to 40 °C
Storage temperature: -10 °C to 50 °C
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa
Height above sea level: Max. 3000 meters
Nominal current draw: Max. 500 mA (CH01 spec)
Power consumption (standby): Max. 2.5 W
Power consumption (charging): Max. 19 W
Approvals: (pending) IEC 60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22.2 No. 60601-1, Australian deviation, Canadian deviation
LED functionality
The charger indicates whether it is connected to mains (green LED) or whether the battery is being charged (orange LED).
Charger mounting/position
The LIFT40 charger CHL40 can be placed for use in two different ways. It is designed for mounting flat on the wall by using minimum 2 of 3 screws in
the Ø4 mm holes in the charger body. Screw types and wall plugs may vary depending on wall material.
The charger can also be placed flat on a table or desk. The dedicated rubber studs on the charger back ensure that the charger stays safely at the
same position during charging (see drawings).
Mounting information
CHL40 is mounted by means of minimum 2 screws (not supplied by LINAK).
Screw type: The screw type depends on the wall type and has to be defined by the service technician.
Washer type: ISO7089 / M5, d1 = 5.3 mm / d2 = 10 mm / s = 1mm.
The CHL40 charger must be mounted with minimum two of the three screws possible.
The mounting screws for the control box and the charger must be tightened with a maximum torque of 1 Nm.
To be able to comply with the IPX4 rating, the CHL40 must hang on the wall.
The charger must be disconnected from mains in cleaning situations.
Mounting holes
Front View
When mounting CHL40 on wall,
please use minimum two of the 3
dedicated mounting holes placed in
body of charger.
Studs for table placement
Studs for table placement
Back View
When placing CHL40 on table of
shelf, 4 rubber studs on the back
adds stability and friction to the
charger / battery combination.
Recommendations
Only use original LINAK components and accessories for full compatibility.
Only use CHL40 charger for BAL40 battery charging.
The charger CHL40 is specifically designed to charge the BAL40 battery.
Special care should be taken when mounting the CHL40.
If the CHL40 is mounted correctly, the CHL40 complies to IPX4.
If the CHL40 is mounted incorrectly, water may enter the screw holes resulting in IPX4 non-compliance and cause malfunction and hazardous situations.
In cleaning situations, the charger must be disconnected from mains.