The HB400 hand control is designed for a wide range of applications such as hospital
beds, nursing home beds and other medical applications such as treatment chairs and
couches.
Usage
Operation temperature: +5 °C to + 40 °C
Storage temperature: -10 °C to + 50 °C
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa
Height above sea level: Max. 3000 meters
Flammability rating: HB
Latex free: Yes
Approvals: IEC 60601-1
ANSI/AAMI ES60601-1
CAN/CSA-22.2 No 60601-1
17. HB400 (MEDLINE
®
CARELINE
®
)
Magnet key - article no. 0858008
Recommendations
It is recommended to make a functional test of the LINAK system in order to ensure that control box and hand control are communicating correctly.
Unless otherwise specified or agreed with LINAK, the hand control is only intended to be used for LINAK systems.
Inform the customer only to use the magnet key supplied by LINAK.
Do not submerge the hand control into water.
It is recommended to check the hand control and cable for damage and holes caused by violent handling before washing the application or at least
once a year.
In order to maintain the cable flexibility, it is important to place a coiled cable in such a way that its own weight does not strain the coil during the
washing process.
When changing hand controls for OpenBus™ systems, the power must be switched off.
When a defective HB400 is replaced, check that the new HB400 has exactly the same specification and functionality.
Warnings
Do not sit or lie on the hand control. It can cause unintended movement of the application.
The application manufacturer must write an end-user manual based on the LINAK user manual which also includes relevant warnings,
information on how to carry out regular inspection and a functionality description. End users must be trained in all functions.
Inform the customer that using the magnet key cannot wake up a low-power system or a system running on battery.
The system will wake up when a key is activated and the magnet key will then unlock the system.
Inform the customer that a powerful magnetic field may change the locking state.
Always use O-ring on connectors and cable locks.