EasyManuals Logo

Linak Inline LA20 User Manual

Linak Inline LA20
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
Page 184 of 295
Recommendations
The customer responsibility includes making a proper design of the cable strain relief inside the side rail panel.
The customer should consider the existence of vibrations when defining and specifying the housing, i.e. we recommend the customer to carry out a
vibration test on the final product.
The customer must ensure a proper IP rating/test
The customer must ensure proper drop testing according to IEC60601-2-52 §201.15.3.4.1. In this clause there is an additional reference to
IEC60601-2-31.
The customer is responsible for correct mounting of the PCBA. Among other things, it means
- ensuring proper and safe mounting of the PCBA into e.g. the side rail
- ensuring proper and correct mounting between key pad connection tails and the ACK PCBA
- ensuring proper and correct mounting of the key pad
- the customer should consider proper precautions against ESD (Electrostatic discharge).
When handling ESDS (Electrostatic Discharge Sensitive) devices – e.g. during transport, storage, handling, production or mounting in an application
- exposure to harmful ESD must be avoided.
Consider proper creepage and clearance measures to fulfil IEC 60601. With One MOPP (One Means Of Patient Protection / Secondary side of the
actuator system)
It is not recommended to dismount the membrane front cover after mounting as this may cause damage.
The ACL (Attendant Control Lock) box is a one turn button box for various applications where the patient
positioning must be carefully controlled by the medical staff.
The ACL disconnects all functions on the hand control either by means of turn button or turn key.
4. ACL (MEDLINE
®
CARELINE
®
)
The Attendant Control OpenBus
TM
(ACO) is a cost optimised and compact unit with up to 21 buttons that can
be used as hand control keys or lock-outs. The lock-out function can be made visable by using yellow LEDs.
Usage
Operation temperature: +5 °C to +40 °C
Storage temperature: -10°C to +50 °C
Relative humidity: 20% to 80% non-condensing
Atmospheric pressure: 700 to 1060 hPa (3000 m)
Flammability rating: V2
Approvals: IEC 60601-1
IEC 60601-1-6
ANSI/AAMI ES60601-1
CAN/CSA-C22.2 NO. 60601-1
In order to comply with the norm, the ACO must hang vertically from its hook during the washing process.
5. ACO (MEDLINE
®
CARELINE
®
)
Recommendations
Always use Locking ring and cables with O-rings.
Locking ring and cables with O-rings must be fitted to ensure IP degree.
If other front covers than standards are requested, the front cover guidelines should be consulted.
N.B.
Cables are inclusive an O-ring.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Linak Inline LA20 and is the answer not in the manual?

Linak Inline LA20 Specifications

General IconGeneral
BrandLinak
ModelInline LA20
CategoryController
LanguageEnglish

Related product manuals