Recommendations
• The control buttons of redundant hand controls for lifting and lowering work as normal hand control buttons.
• Please be aware that loss of power might occur due to the battery deep discharge protection.
This will only be the case if the battery is continually being used despite a warning.
• The service counter is only active in CAL40+ when a sufficiently charged BAL40 is mounted.
• The battery pack BAL40 must not be removed in cleaning situations, doing so could result in non-compliance with IPX6.
• If LIFT40 is fitted with a mains power connector, the protection plug must always be inserted to ensure the IP protection, if the port is not used.
• Only use original LINAK mains cables to ensure a proper connection to the internal charger.
• When charging, LIFT40 will not be able to operate any actuators.
• The LIFT40 DC plug is intended for charging of the BAL40 battery.
Using the DC plug for powering external equipment can lead to battery drainage or discharge.
• Only use correct LINAK charger (CHL40, CH01 or integrated charger in LIFT40).
• When customising CAL40 software via LIXedit tool, be sure to follow LINAK actuator parameters for current settings
to ensure correct compatibility of actuators and CAL40.
Warning
• Never connect the programming box directly to the hand control port.
• To avoid injury, the battery should not be mounted in transport situations. Use LINAK original packaging to store battery during transportation.
• Use blind plug when cleaning/washing down to maintain the IP degree.
• When using a control box with emergency stop, the stop button must be activated in cleaning situations to avoid unintended operation of the lift.
• In order to avoid injury, the emergency stop should be activated in all transport and cleaning situations.