Operating mode Fault Cause Remedy
Battery
operation
No indication Battery empty Change battery
Battery contacts
corroded
Clean contacts
After insertion of
the battery the
Lutz company logo
appears longer than 1
minute
Basicowmeteris
missing
Checkbasicowmeter
Check connecting wires and
plug connections
No indication after
changing batteries
Batteriestted
incorrectly
Open battery compartment
and check polarity
Battery contacts
corroded
Clean contacts
Wrong battery type
used
Battery type Operating
instructions operating unit
chapter 10.2
Operation with
relay module
Buttons RESET, START,
Change volume, Cali-
bration, Total volume,
Flow monitoring,
MeasurementOn/O,
Coupling modules,
Factory settings are
not available
Basicowmeteris
missing
Checkbasicowmeter
Check connecting wires and
plug connections
Buttons START,
Change volume,
Economy, Backlight,
Safety prompt are not
available
Operating unit is
powered by battery
and relay module is
missing
Check mains connection
Check fuses in the relay
module
Check connecting wires and
plug connections
In case of operation with relay
module, batteries are not
necessary.
74