3.1.2 Calibrating a probe using a ledge
When calibrating a disc probe it is advisable to calibrate this disc
probe with a ledge
Press the „Calibrate a probe“ key
Position the probe below the ledge
Press the „Calibrate a probe using a ledge“ key
The probe makes contact
Probe travels automatically upwards
Position the ledge below the probe
The probe makes contact
Note: Procedure will be repeated automatically
Probe travels automatically downwards
Position the probe a second time below the ledge
The probe makes contact
Probe travels a second time above the ledge
Position the ledge a second time below the probe
The probe makes contact
Each time the probe makes contact an acoustic signal (beep) will be
emitted
Once the calibration has been successful, the determined probe
constant will appear in the display.