3.1.2 Calibrating a probe using a ledge
When calibrating a disc probe it is advisable to calibrate this disc
probe with a ledge
– Press the „Calibrate a probe“ key
– Position the probe below the ledge
– Press the „Calibrate a probe using a ledge“ key
– The probe makes contact
– Probe travels automatically upwards
– Position the ledge below the probe
– The probe makes contact
Note: Procedure will be repeated automatically
– Probe travels automatically downwards
– Position the probe a second time below the ledge
– The probe makes contact
– Probe travels a second time above the ledge
– Position the ledge a second time below the probe
– The probe makes contact
Each time the probe makes contact an acoustic signal (beep) will be
emitted
Once the calibration has been successful, the determined probe
constant will appear in the display.