Description / Sequence Symbols / Pictures
8.1.3 Exchanging the batteries
The battery can be changed without the height measuring instrument
losing data (except for the time and date).
Switch OFF the height measuring instrument
Disconnect the charging device from the height measuring
instrument
Remove the battery cover, by unscrewing the 2 knurled thumb
screws, ill. 1
Carefully remove the battery pack out off the retaining spring
(clips), ill. 2
Press the new battery pack into the retaining spring and reconnect
the plug to the battery pack, ill. 4
Replace the battery cover and insert and tighten the knurled
thumb screws
Attach the new battery pack to the mains adapter and charger for
a min. 5 hours.
Attention
Please only use the specified battery pack!
Order no. 4429449 - NI-MH 4.8 V - 7000mAh (with 3 cables)