Software User’s Guide
Default Settings
Load and Distance Units
Select the desired units of measure. Note that not all units are supported by all force sensors. Refer to the
force sensor’s specifications for list of supported units.
Digital Filter
The digital filter smooths out load readings in situations where there is mechanical interference in the
work area or test sample. The filter utilizes a moving average technique in which consecutive readings
are pushed through a buffer and the displayed reading is the average of the buffer contents. By varying
the length of the buffer, a variable smoothing effect can be achieved. The selection of 1 will disable the
filter since the average of a single value is the value itself.
Available settings: 1 through 16,384.
It is recommended to keep the filter as low as possible for best performance.
Default Folders
Select the corresponding Open buttons to browse to the desired folder locations for test configuration
files, data, results, and reports.
All above file types created by IntelliMESUR are encrypted, for data integrity purposes, except .csv file
exports.