Series F Test Frames + IntelliMESUR
®
Software User’s Guide
68
TOC
12.1.2 Slider lubrication once per year
Slider lubrication frequency should increase when used in in adverse conditions. Follow these
instructions:
Two people are recommended for this procedure, due to the risk of tipping.
1. Flip the test frame upside down, as shown below. Be sure to do this on a flat and secure working
surface.
2. While one person is supporting the column, the other person removes six screws from the
underside of the base of the test frame, as shown below: