Table 31 Alarm 1 and Alarm 2 Maintained Alarm Logic
ALARM AL1
MAINTAINED ALARM
(when in abnormal condition ack is pressed)
CONDITION
NORMAL
ABNORMAL
UP (O/S)
DOWN (O/S)
ACK
(2)
NORMAL
(1)
ACK
(3)
ALARM
LATCH
LAMP
OFF
FLASH
FLASH
OFF
STEADY
STEADY
OFF
YES
RELAY / OPEN
COLLECTOR
OFF
ON
ON
OFF
OFF
OFF
OFF
HIGH
ALARM
LATCH
LAMP
OFF
FLASH
FLASH
OFF
STEADY
OFF
OFF
SP1
NO
RELAY / OPEN
COLLECTOR
OFF
ON
ON
OFF
OFF
OFF
OFF
TRIP
LAMP
OFF
FLASH
OFF
OFF
STEADY
STEADY
OFF
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
OFF
ON
ON
OFF
ALARM
LATCH
LAMP
OFF
FLASH
OFF
FLASH
STEADY
STEADY
OFF
YES
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
ON
OFF
OFF
OFF
LOW
ALARM
LATCH
LAMP
OFF
FLASH
OFF
FLASH
STEADY
OFF
OFF
SP1
NO
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
ON
OFF
OFF
OFF
TRIP
LAMP
OFF
FLASH
OFF
OFF
STEADY
STEADY
OFF
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
OFF
ON
ON
OFF
ALARM AL2
MAINTAINED ALARM
(when in abnormal condition ack is pressed)
CONDITION
NORMAL
ABNORMAL
UP (O/S)
DOWN (O/S)
ACK
(2)
NORMAL
(1)
ACK
(3)
ALARM
LATCH
LAMP
OFF
FLASH
FLASH
OFF
STEADY
STEADY
OFF
YES
RELAY / OPEN
COLLECTOR
OFF
ON
ON
OFF
OFF
OFF
OFF
HIGH
ALARM
LATCH
LAMP
OFF
FLASH
FLASH
OFF
STEADY
OFF
OFF
SP2
NO
RELAY / OPEN
COLLECTOR
OFF
ON
ON
OFF
OFF
OFF
OFF
TRIP
LAMP
OFF
FLASH
OFF
OFF
STEADY
STEADY
OFF
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
OFF
ON
ON
OFF
ALARM
LATCH
LAMP
OFF
FLASH
OFF
FLASH
STEADY
STEADY
OFF
YES
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
ON
OFF
OFF
OFF
LOW
ALARM
LATCH
LAMP
OFF
FLASH
OFF
FLASH
STEADY
OFF
OFF
SP2
NO
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
ON
OFF
OFF
OFF
TRIP
LAMP
OFF
FLASH
OFF
OFF
STEADY
STEADY
OFF
RELAY / OPEN
COLLECTOR
OFF
ON
OFF
OFF
ON
ON
OFF
Notes :
(1)
means normal condition after abnormal has occurred
(2)
means ack pressed in abnormal condition
(3)
means ack pressed in normal condition after abnormal has already occurred.