The instrument employs the electronic impedance principle for cell counting and
sizing, and a colorimetric method for measuring hemoglobin. A microprocessor
is used to measure the parameters and to size the cells. During the count the pro-
cessor is checking the analysing process for any irregularities. Size distributions
are printed for all populations (RBC, PLT and WBC).
The instrument has as standard a parallel and a serial output, which is user pro-
grammable. From a built-in selectable program the user can choose between dif-
ferent print formats.
A bar-code reader (scanner), available as an option, can be connected to enter the
sample ID automatically.
The instrument is fully automatic, which means that, the system is always pow-
ered on and performs automatically preforms check- and cleaning cycles to min-
imize user maintenance.
Note:
PCT, LPCR, RDWa and PDWa are not established parameters. Their use should
be restricted to research or investigational use only.
2.3 Short-List of Specifications
The Platelet Distribution Width (Absolute) (PDW)
The Platelet Macro Range > 12 fl in% (LPCR)
The Platelet Packed Volume (PCT)
Measuring principle
RBC, WBC, PLT
Impedance
Measuring principle HGB Cyanide free method 540nm
Discriminator Floating programmable
Sampling system Closed shear valve
Parameters reported RBC, MCV, HCT, PLT, MPV HGB, MCH,
MCHC, WBC,
RDW%, LYMF abs, MID abs, GRAN
abs., LYMF%, MID%, GRAN%
RDW abs, PDW abs, LPCR, PCT
Size distributions printed for RBC, PLT and WBC diff.
Aspirated blood volume
(open tubes)
approx. 125µl
Aspirated blood volume
(closed tubes)
1
approx. 200 µl
Blood volume using the
Micro Pipette Adapter
20 µl
Pre-diluted mode 1:200 to 1:250 using min. 20µl blood
e.g. 20 µl to 5 µl diluent
30 µl to 6 µl diluent
40 µl to 8 µl diluent
CA530 Screen 2x40 character LCD
CA620 Screen Graphic LCD display
Keyboard Numerical
Total cycle time approx. 73 seconds
Important
Applicable for CA620-
CellGuard:
The MPA is intended to be
used for pre-donor screen-
ing. Using the MPA for
PLT concentrate may re-
duce the accuracy of the
result.