2.7 Differences between CA620 and
CA530 (9,16,20 parameters)
The CA620 and CA530 have identical parameter performances in all respects.
The differences are the following:
• -The INTRA mode, see section Analysing the Sample (Open Tube) on
page 54 and “Block parameters” see section Block parameters on page 72
are not available on the CA530.
• The CA620 has an extended calibration menu, see Calibration Through a
Certified Calibrator on page 83.
• The CA620 has an extended QC function including X-B and Levey-Jen-
nings Plots. See Introduction on page 87.
• The CA530 employs a 2 x 40 alphanumeric display and the CA620 is
equipped with a graphic display making, a graphic display of the size-distri-
bution curves, possible.
• Extended print format is available on the CA620 only.
See appendix 530-30-205 available from your authorized distributor.
• The CA620-CellGuard can be configured as a donor screening device or as
a tool to determine the PLT concentrations in PLT bags. In case the instru-
ment is used for donor screening, it is fully identical to a CA620 16 param-
eter machine.
• The CA620 differs from the CA530 by the specially designed front panel
and keyboard.