EasyManua.ls Logo

Medonic CA620 - Hgb

Medonic CA620
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration and controls
84 03-11-24 1009en01
The following calibration procedure is used for the CA620 and the CA620-Cell-
Guard, using a Boule certified whole blood calibrator:
1. Run the calibrator 5 times at the open tube inlet. Follow the guidelines
shown in Use of Calibrators and Controls on page 81 above. Enter the
ID 888+ for each run; entered as 888 and terminated with the right-arrow
key. See Print control blood id:s on page 70. In case the calibrator was read
by a bar-code reader as outlined in Initializing the Levey-Jennings Plots
and Functions on page 87, use the calibrator ID instead of the 888+ entry.
To display a quick overview which controls and calibrators are defined, go
to menu 5.10.9 and press [Enter] to print a list with control/calibrator def-
initions.
2. Go to menu 5.1. The following is displayed:
The mean value and CV of the last 5 samples are shown.
3. Scroll with the “left/right” key to the parameter that has to be calibrated and
check that the shown CV values
1
are less than:
In case one run should be excluded from the auto calibration to reduce the CV,
move the cursor to the specific sample and press [Enter]. The display of “Use”
will show “0” on such sample, indicating that it is not included in the auto cali-
bration.
It is also possible to print out a sample by moving the cursor to a sample line and
pressing the [Print] button.
WBC < 4%
RBC < 1.8%
PLT < 6%
MCV < 1%
HGB < 1.8%
MPV < 5% (if applicable)
1. The CV values are stated higher than the typical CV of the instrument due to
the limited number of sample runs.
1139en.gif

Table of Contents

Related product manuals