4 Parameter Flags
The instrument has several indication- and warning flags related to the measured
parameters. The flags are shown on the display as well as printed on the connect-
ed printer. An error flag should be seen as parameter value error. The sample
should be re-analyzed. A warning flag might indicate a pathological sample. If on
the same parameter both an error and a warning flag occurs; the error flag has
always a priority and is displayed and printed in the first place. Note that a param-
eter that is outside the “Normal Range” which is set in menu 5.4 is either marked
with “H” or “L” on the connected printer to indicate if the value is Higher or
Lower than the pre-set “Normal Range” values. The display will in such a case
show a “*” in front of the parameter value.
The following flags are seen as “Error flags” or “Warning flags”
Flag Cause Action Service call Comment
TU Blockage in aperture Re-run sample If persist See TU (Time-out Up-
per Detector) on page
34.
TL Blockage in aperture Re-run sample If persist See TL (Time-out Low-
er Detector) on page 34.
SE Irregular count Re-run sample See SE (Statistical Er-
ror) on page 34.
DE Interference between 2
populations
Examine size distribution
RBC/PLT or WBC
See DE (Distribution
Error) on page 34.
FD Insufficient cell
separation RBC/PLT
Examine size distribution
RBC/PLT
See FD (Floating Dis-
criminator) on page 35.
OF HGB photometer
offset out of range.
Wait 15-30 minutes and
restart instrument.
If persist See OF (Offset error
HGB) on page 35.
LO HGB photometer lamp
intensity too low
See comment If persist See LO (Low blank lev-
el HGB) on page 35.
HI HGB photometer lamp
intensity too high
See comment If persist See HI (High blank lev-
el HGB) on page 35.
NG Negative HGB value Run “Prime” from menu
8.1
If persist See NG (Negative
HGB) on page 35.
SE irregular HGB
measurement
Re-run sample If persists on all sam-
ples
See SE (Statistical error
HGB) on page 36.
TB Air bubbles in system Run ‘Prime’ from menu
8.1
If persist See TB (Air bubbles) on
page 36.
NM,
OM,
TM,
BD
WBC differential abnor-
mality
Examine sample by refer-
ence method
See NM, OM, TM, BD
(WBC Differential
Flags) on page 36.
RP Red cells present in PLT
concentrate (CA620-Cell-
Guard only)
Disregard PLT
parameter result
See RP (Red Cells
Present) on page 36.
EC Blood control expired Use new blood
control
See EC (Expired con-
trol) on page 37.