EasyManua.ls Logo

Medonic CA620 - Analyzing the Sample (Pre-Dilute)

Medonic CA620
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Routine Operation
56 03-11-24 1005en01
Sample displayed after ca. 53 seconds from aspiration:
7.3 Analyzing the Sample (Pre-Dilute)
Any dilution rate between 1:200 and 1:250 can be used under the condition that
the minimum total volume is 5ml and the maximum volume 8 ml. It is obvious
that always the same dilution rate must be chosen as any reproducibility error in
an external dilution will directly affect the counted parameters.
Examples:
20 µl and 5 ml diluent, 30 µl and 6 ml diluent or 40 µl and 8 ml diluent
The sample should be analyzed as soon as possible. Prolonged waiting increases
the inaccuracy of the MCV and WBC differential parameters.
Place the cup with the pre-diluted sample under the aspiration pipette for
pre-diluted samples and press the lever behind the pre-dilute aspiration pi-
pette.
Note that the instrument aspirates more than 5 ml. Volumes over 5ml are not
specified. This means that in most cases the whole sample (volume) is aspirated
into the analyzer.
Remove the sample cup after aspiration.
The display sequence on the LCD screen is identical to the description in Rou-
tine Operation on page 53 above.
If no sample was entered and the pre-dilute start lever was pressed anyhow, the
system will recognise this and restore the liquid flow system automatically.
1059en.gif
Important
The instrument is ready to
accept the next sample af-
ter the flashing display cur-
sor in the upper left display
corner disappears.
Entering the next sample
before the flashing cursor
disappears will result in er-
roneus parameter results
for the next sample.
1060en.gif

Table of Contents

Related product manuals