EasyManua.ls Logo

Medonic CA620 - Filling the System with Reagents

Medonic CA620
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation
46 03-11-24 1003en01
Note:
Aspiration command keys are not located on the keyboard. Aspiration of a sam-
ple is started by pressing the start-lever behind the aspiration needle.
5.5 Filling the system with reagents
After power-on, the instrument will perform a self-check which lasts for ca. 2
minutes. During this period, no key entry is accepted by the instrument.
1. Press
[Menu]until the main menu is displayed on the LCD screen.
2. Scroll with the up-arrow key to 8 and press
[Enter].
3. Scroll to 8.2 and press [Enter].
The system is now filled with reagents. This cycle lasts for ca. 6 minutes.
When the system is finished with this cycle, the cursor field returns to the field
8.2. Press
[Menu] to return to the main menu.
1037en.gif1036en.gif

Table of Contents

Related product manuals