2.5 Reagents and Reagent Consumption
The reagent consumption is depending on the total number of samples/day. Us-
ing the system for more than ca. 50 samples/day; the reagent consumption will
be:
The approximate consumption can be calculated from the graph below and is de-
pending on the number of samples/day. The figures stated are indicative only and
assumes maximum 2 “blank” counts per day. As seen from the graph, the relation
Diluent/Lyzer is 2:1 in case of 50 samples/day or more. Hence, using the instru-
ment for less number of samples/day, less Diluent is consumed in relation to the
lyzer
Diluent Consumption
Lyzer consumption
For additional information regarding cleaning solutions; please refer to Mainte-
nance, Shut-Down & Transport on page 95.
Diluent approx. 19 ml/sample
Lyzer approx. 9.5 ml/sample
Caution
Use only by Boule autho-
rized reagents.
Erroneous results and
damage of the system may
occur if other reagents are
used.
Diluent consumption CA620 vs # samples
0
5
10
15
20
25
30
35
0 50 100 150 200
Number of Blood_samples/Day
Diluent consumption / sample
(ml)
1014.eps
Lyzer consumption CA620 vs # Samples
0
2
4
6
8
10
12
14
16
18
20
0 50 100 150 200
Number of Blood_samples/Day
Lyzer consumption /
sample (ml)
1015.eps