2.4 Parameter Ranges
Linearity +/-1%
within the following range:
WBC 0.5 - 80.0
RBC 0.5 - 9.99
MCV 55 - 130
PLT 30 - 999
HGB 0.5 - 55.0
PLT-Concentrate mode
(RBC < 0.3 and measured on a Latex
solution)
30-6999
(applicable on CA620-Cellguard only)
Measuring range:
WBC 0 - 99.9
RBC (CA620-Cellguard) 0 - 13.99
RBC (CA620/530) 0 - 13.99
MCV 15 - 250
PLT 0 - 6999
HGB 0 - 99.9
PLT Concentrate mode 0-6999
(applicable on CA620-Cellguard only)
Correlation
1
1. Using Coulter AcT and Abbot CD1600 as reference
RBC, WBC R > 0.97
PLT R > 0.90
HGB R > 0.99
MCV R > 0.90
Reproducibility (typical):
Measured as an average of 10 measurements each of 10 different vein K2-
EDTA collected normal samples:
Parameter X-mean (CGS units) CV(%)
WBC 7.4 2.0
RBC 4.52 0.85
MCV 89.4 0.5
PLT 252 3.2
HGB 13.9 0.8