Routine Operation
1005en01 03-11-24 59
2. Push (rotate) the protection cover to the left and insert the tube up-side
down.
3. Push (rotate) the protection cover to the right.
The aspiration cycle is now started following the sequence as described in Anal-
ysing the Sample (Open Tube) on page 54.
Remove the tube after the sample parameters are displayed in reversed order. The
adapter is ready to accept the next tube when the sample parameters are displayed
and the flashing cursor in the parameter-field of the display disappears.
Caution
Insert the sample tube
with lid facing downwards.
Ignoring this instruction
may damage the aspiration
needle.
1067.bmp
Caution
Avoid using the same tube
more than twice in the cap
piercing device.
Not doing so might in-
crease the risk that rubber
particles from the lid clogs
the instrument which leads
to incorrect results and/or
excessive service.
1068.bmp