3.9 Connecting the instrument to the power grid
■■■■■■■■■■■■■■■■■■■■■■
16
■■■■■■■■
876 Dosimat plus
3.9 Connecting the instrument to the power grid
WARNING
Electric shock from electrical potential
Risk of injury by touching live components or through moisture on live
parts.
■ Never open the housing of the instrument while the power cord is
still connected.
■ Protect live parts (e.g. power supply unit, power cord, connection
sockets) against moisture.
■ Unplug the power plug immediately if you suspect that moisture has
gotten inside the instrument.
■ Only personnel who have been issued Metrohm qualifications may
perform service and repair work on electrical and electronic parts.
Connecting the power cord
Accessories
Power cord with the following specifications:
■ Length: max. 2 m
■ Number of cores: 3, with protective conductor
■ Instrument plug: IEC 60320 type C13
■ Conductor cross-section 3x min. 0.75 mm
2
/ 18 AWG
■ Power plug:
– according to customer requirement (6.2122.XX0)
– min. 10 A
NOTICE
Do not use a not permitted power cord!
1
Plugging in the power cord
■ Plug the power cord into the instrument's power socket.
■ Connect the power cord to the power grid.