5 System settings
876 Dosimat plus
■■■■■■■■
35
Selection on | off
Default value off
Time interval
This parameter is visible only when Monitoring = on.
You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.
Input range 1 - 999 d
Default value 999 d
5.3 File management
Menu System File management
NOTICE
This menu item is visible only if a USB flash drive has been connected as
an external storage medium.
Methods can be imported and deleted from a USB flash drive in this dia-
log. Only methods located in the Files directory are displayed in the list
(see "Directory structure on the USB flash drive", page 36).
A backup of the system can be created (all data and settings). Similarly, an
existing backup can be reloaded.
Import
Import the selected method.
Delete
Delete the selected method.
Backup
Create a backup of all data and settings on the USB flash drive.
NOTICE
Only one backup can be created on the same USB flash drive.
If a backup is already stored on the flash drive, then this will be over-
written as soon as the function is performed once again.
Restore
Load the backup from a connected USB flash drive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm Dosimat plus 876 and is the answer not in the manual?

Metrohm Dosimat plus 876 Specifications

General IconGeneral
BrandMetrohm
ModelDosimat plus 876
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals