5 System settings
876 Dosimat plus
■■■■■■■■
35
Selection on | off
Default value off
Time interval
This parameter is visible only when Monitoring = on.
You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.
Input range 1 - 999 d
Default value 999 d
5.3 File management
Menu ▶ System ▶ File management
NOTICE
This menu item is visible only if a USB flash drive has been connected as
an external storage medium.
Methods can be imported and deleted from a USB flash drive in this dia-
log. Only methods located in the Files directory are displayed in the list
(see "Directory structure on the USB flash drive", page 36).
A backup of the system can be created (all data and settings). Similarly, an
existing backup can be reloaded.
Import
Import the selected method.
Delete
Delete the selected method.
Backup
Create a backup of all data and settings on the USB flash drive.
NOTICE
Only one backup can be created on the same USB flash drive.
If a backup is already stored on the flash drive, then this will be over-
written as soon as the function is performed once again.
Restore
Load the backup from a connected USB flash drive.