5.2 Managing solutions
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
876 Dosimat plus
Cylinder volume
Cylinder volume of the buret unit in mL. The cylinder volume is automati-
cally read out with intelligent buret units.
Selection 1 | 5 | 10 | 20 | 50
Default value 20
Concentration
Concentration of the solution.
Input range –999999999 - 9999999999
Default value 1.000
Concentration unit
Unit of the concentration.
Selection µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L | User-defined
Default value mol/L
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Titer
Titer of the solution.
Input range –999999999 - 9999999999
Default value 1.000
Titer unit
Unit of the titer.
Selection µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L | empty | User-defined
Default value empty
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Date titer det.
Date of the last titer determination.
Monitoring
Activating and deactivating the titer monitoring.