6.1 Manual dosing (DOS)
■■■■■■■■■■■■■■■■■■■■■■
42
■■■■■■■■
876 Dosimat plus
report if the corresponding settings have been carried out, see the follow-
ing Chapter.
The result of the calculation is shown under the volume display. The result
and the dosing volume can be deleted by pressing the [FILL] key once
more.
6.1.2 Selecting a solution
Menu ▶ Parameters
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. Solutions are defined under System ▶ Solutions.
For exchange units with integrated data chip, a check is made in the
method run to verify whether the correct solution has been attached. For
exchange units without integrated data chip, only the cylinder volume is
checked. The validity of the titer is checked for the selected solution at the
start of the determination.
Selection Selection of configured solutions | not defined
Default value not defined
not defined
No check takes place.
6.1.3 Dosing parameters
Menu ▶ Parameters ▶ Dosing parameters
Dosing rate
The rate at which dosing takes place. The maximum dosing rate depends
on the cylinder volume (see chapter 8.1.1, page 53).
Input range 0.01 - 150.00 mL/min
Selection max.
Default value max.
Filling rate
Rate at which the dosing cylinder is to be refilled. The maximum filling rate
depends on the cylinder volume (see chapter 8.1.1, page 53).
Input range 0.01 - 150.00 mL/min
Selection max.
Default value max.
Mode
Method of dosing.