6.2 Extended dosing (XDOS)
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
876 Dosimat plus
Volume
The volume to be dosed.
Input range 0.0000 - 99999.9 mL
Default value 10.0000 mL
Dosing rate
The rate at which dosing takes place. The maximum dosing rate depends
on the cylinder volume (see chapter 8.1.1, page 53).
Input range 0.01 - 150.00 mL/min
Selection max.
Default value max.
Filling rate
Rate at which the dosing cylinder is to be refilled. The maximum filling rate
depends on the cylinder volume (see chapter 8.1.1, page 53).
Input range 0.01 - 150.00 mL/min
Selection max.
Default value max.
Volume limit
Safety limit for limiting the maximum volume to be dosed.
Input range 0.01 - 99999.9 mL
Selection off
Default value off
off
No safety limit.
Auto fill
Automatic filling of the cylinder after the dosing.
Selection off | on
Default value on
6.2.5 Reports
Menu ▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated result and additional specifica-
tions.