■■■■■■■■■■■■■■■■■■■■■■
6 Parameters and dosing modes
876 Dosimat plus
■■■■■■■■
45
Delete the unit entered under User-defined. A blank entry will be gener-
ated in the selection list.
Selection % | /pc | L | g | g/L | mL | mg | mg/mL | mol |
mol/L | ppm | User-defined
Default value %
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.
Factor
Factor for the calculation formula.
Entry 10 digits
Default value 1.0
Divisor
Divisor for the calculation formula.
Entry 10 digits
Default value 1.0
Blank
The blank value is subtracted from the dosing volume.
Input range 0.0000 - 1,000,000 mL
Default value 0.0000 mL
6.1.5 Reports
Menu ▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated result and additional specifica-
tions.
Selection off | on
Default value off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 36).