4 Operation
876 Dosimat plus
■■■■■■■■
27
4.6 Preparing the buret unit (PREP)
The PREP function is used to rinse the cylinder and tubings of the buret
unit and fill them air bubble-free. You should carry out this function once
per day.
Proceed as follows:
1
■ In the main dialog, select Menu and press [OK].
The main menu opens.
■ Select the menu item Prepare buret unit and press [OK].
If an 805 Dosimat for tandem dosing is connected to the instru-
ment, select the first or second Dosimat afterwards.
The following message is displayed:
2
Start the preparing
CAUTION
Make sure that the buret tip is directed into a vessel that can
accommodate the cylinder volume of your buret unit several times
over.
■ Select Yes and confirm the message with [OK].
Preparing is carried out.