6.2 Extended dosing (XDOS)
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
876 Dosimat plus
6.2.2 Selecting a solution
Menu ▶ Parameters
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. Solutions are defined under System ▶ Solutions.
For exchange units with integrated data chip, a check is made in the
method run to verify whether the correct solution has been attached. For
exchange units without integrated data chip, only the cylinder volume is
checked. The validity of the titer is checked for the selected solution at the
start of the determination.
Selection Selection of configured solutions | not defined
Default value not defined
not defined
No check takes place.
6.2.3 Selecting the operating mode
Menu ▶ Parameters
Buret setup
If continuous dosing without interruption is required, then the tandem
mode can be selected. An additional 805 Dosimat is required for this pur-
pose.
Selection Single | Tandem
Default value Single
Single
Dosing with one instrument.
Tandem
Dosing with two Dosimats.
Tandem operation
An 805 Dosimat is connected to the MSB connector of the 876 Dosimat
plus for tandem operation.
NOTICE
Two exchange units with the same cylinder size must be used for tan-
dem operation.
The principle of tandem dosing: