5.1 Basic settings
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
876 Dosimat plus
Routine
Limited dialog for routine operations.
Contrast
The contrast of the display can be adjusted with the arrow keys [] and
[].
[]: the contrast will be decreased by one step each time the key is
pressed.
[]: the contrast will be increased by one step each time the key is
pressed.
Input range 150 - 240
Default value 212
NOTICE
Alternatively, the contrast can also be modified in the following man-
ner:
Keep the red [STOP] key pressed down. As soon as the progress bar
appears, also press the arrow key [] or [] repeatedly.
This method will, however, cause the contrast to be modified by several
steps.
Beep
If this parameter is activated, then a short beep will sound in the following
cases:
When a key is pressed.
At the end of the determination.
Selection on | off
Default value on
PREP warning
If this parameter is activated, then the recommendation will be made to
carry out the function PREP (Preparing):
After the instrument is switched on.
Each time a buret unit is attached.
All tubing and the cylinder are rinsed with this function (see chapter 8.1.2,
page 53).
Selection on | off
Default value on