5.4 Configuring external devices
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
876 Dosimat plus
The interface parameters for the connected balance are set under COM1
settings.
Baud rate
Transfer rate in characters per second.
Selection 1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200
Default value 9600
Data bits
Number of data bits.
Selection 7 | 8
Default value 8
Stop bits
Number of stop bits.
Selection 1 | 2
Default value 1
Parity
Type of parity testing.
Selection even | none | odd
Default value none
Handshake
Type of the data transfer protocol.
Selection hardware | software | none
Default value hardware
NOTICE
If communications problems occur, set the parameter Handshake to
software, and make another attempt.
Editing the COM2 settings
Menu ▶ System ▶ External devices ▶ COM2 settings
The interface parameters for instruments connected to the RS-232/2 con-
nector of the RS-232/USB Box (e.g. PC) are set under COM2 settings.
The parameters and input ranges are identical to those for the COM1
interface.