Page #71 background imageLoading...
Page #71 background image
66
<Command> :MEASure:VOLTage:AC <voltage NR3/disc>
:MEASure:VOLTage:AC?
FunctionIt sets or queries the AC measurement voltage (LEVEL).
Description
Set parameterWhen RO(output impedance) is setting 100
Ω
, the setup value of AC voltage is
{0.01~2|
MAXimum|MINimum}.When RO(output impedance) is setting 25Ω, the setup value
of AC voltage is {0.01~1|MAXimum|MINimum}.
Set syntax
:MEASure:VOLTage:AC 1
:MEASure:VOLTage:AC 1M
:MEASure:VOLTage:AC 1MV
:MEASure:VOLTage:AC 1E-3
:MEASure:VOLTage:AC MAXimum
:MEASure:VOLTage:AC MINimum
Query syntax
:MEASure: VOLTage:AC?
Return data1.000000E+00 (Format is in <NR3>)
When the Level setup is current mode, the data returns 9.9E37.
<Command> :MEASure:VOLTage:DC <voltage NR3/disc>
:MEASure:VOLTage:DC?
FunctionIt sets or queries the DC measurement voltage (LEVEL).
Description
Set parameterThe value of DC voltageis {0.01~1|
MAXimum|MINimum}.
Set syntax
:MEASure:VOLTage:DC 1
:MEASure:VOLTage:DC 1M
:MEASure:VOLTage:DC 1MV
:MEASure:VOLTage:DC 1E-3
:MEASure:VOLTage:DC MAXimum
:MEASure:VOLTage:DC MINimum
Query syntax
:MEASure:VOLTage:DC?
Return data 1.000000E+00 (Format is in <NR3>)
When the Level setup is current mode, the data returns 9.9E37.
<Command> :MEASure:CURRent:AC <current NR3/disc>
:MEASure:CURRent:AC?
FunctionIt sets or queries the AC measurement current (LEVEL).
Description
Set parameterWhen RO(output impedance) is setting 100Ω, the setup value of AC current is
{0.0002~0.02|MAXimum|MINimum}. When RO(output impedance) is setting 25Ω, the setup
value of AC current is {0.0002~0.04|MAXimum|MINimum}.
Set syntax
:MEASure:CURRent:AC 0.01
:MEASure:CURRent:AC 10M
:MEASure:CURRent:AC 10MA
:MEASure:CURRent:AC 1E-2
:MEASure:CURRent:AC MAXimum
:MEASure:CURRent:AC MINimum
Query syntax
:MEASure:CURRent:AC?
Return data1.000000E-02 (Format is in <NR3>)
When the Level setup is voltage mode, the data returns 9.9E37.

Table of Contents