<Command> :MEASure:PARAmeter
{OFF|RDC|LS|LP|CS|CP|Q|D|RS|RP|Z|DEG|RAD|R|X|Y|G|B|E|U}
:MEASure:PARAmeter?
FunctionIt sets or queries the measurement parameter (max.4 items) at present.
Description
Set parameter OFF, RDC(DC Resistance), Ls(Series Inductance), Lp(Parallel Inductance),
Cs (Series Capacitance), Cp (Parallel Capacitance), Q (Quality Factor), D (Dissipation
Factor), Rs (Series Resistance) Rp(Parallel Resistance), Z(Impedance), θd(Angle),
θr(Diameter), R(Resistance), X(Reactance), Y (Admittance), G (Conductivity), B
(Susceptance), E(Relative Permittivity), U(Relative permeability)
Set syntax
:MEASure:PARAmeter RDC,Z,DEG,OFF
RDC is the first measure parameter, the second is Z, the third is θ
deg
and the forth is OFF.
Query syntax
:MEASure:PARAmeter?
Return dataRDC,Z,DEG,OFF
<Command>
:MEASure:FREQuency<frequencyNR3/disc>
:MEASure:FREQuency?
FunctionIt sets or queries the FREQUENCY.
Description
Set parameter
The setup value of frequency is {10.0~30000000.0|MAXimum|MINimum}.
Set syntax
:MEASure:FREQuency 1000
:MEASure:FREQuency1K
:MEASure:FREQuency 1KHZ
:MEASure:FREQuency 1E3
:MEASure:FREQuency MAXimum
:MEASure:FREQuency MINimum
Query syntax
:MEASure:FREQuency?
Return data 1.000000E+03 (Format is in <NR3>)
<Command> :MEASure:SPEEd
{MAXimum|FAST|MEDium|SLOW|SLOW2|0|1|2|3|4}
:MEASure:SPEEd?
FunctionIt sets or queries the SPEED.
Description
Set parameter MAXimum/0, FAST/1, MEDium/2, SLOW/3, SLOW2/4
Set syntax
:MEASure:SPEEd 1
:MEASure:SPEEd FAST
Query syntax
:MEASure:SPEEd?
Return data FAST (Format is in <disc>)
<Command> :MEASure:BEEPer {OFF|PASS|OK|FAIL|NG|0|1|2]
:MEASure:BEEPer?
FunctionIt sets or queries the BEEP WHEN.
Description
Set parameter
OFF, PASS|OK, FAIL|NG
Set syntax
:MEASure:BEEPer PASS
:MEASure:BEEPer NG
:MEASure:BEEPer OFF
Query syntax
:MEASure:BEEPer?
Return dataPASS| FAIL| OFF