73
LIST Subsystem
<Command>
:LIST:STEP{1|2|3|4|…|14|15}
:LIST:STEP?
FunctionIt sets or queries the step of list mode.
Description
Set parameter
step number1~15
Set syntax
:LIST:STEP 1
Query syntax
:LIST:STEP?
Return
data1(Format is in <NR1>)
<Command>
:LIST:PARAmeter
{OFF|RDC|LS|LP|CS|CP|Q|D|RS|RP|Z|DEG|RAD|R|X|Y|G|B}
:LIST:PARAmeter?
FunctionIt sets or queries the measurement parameter of list mode.
Description
Set parameterThere is no E(Relative Permittivity) and U(Relative permeability) in the list
mode.
Set syntax
:LIST:PARAmeter Z
Query syntax
:LIST:PARAmeter?
Return
dataZ(Format is in <disc>)
<Command>
:LIST:FREQuency <frequencyNR3/disc>
:LIST:FREQuency?
FunctionIt sets or queries the measurement frequency of list mode.
Description
Set parameter
frequency 10.0~30000000.0MAXimum/MINimum
Set syntax
:LIST:FREQuency 1000
:LIST:FREQuency 1K
:LIST:FREQuency 1KHZ
:LIST:FREQuency 1E3
:LIST:FREQuency MAXimum
:LIST:FREQuency MINimum
Query syntax
:LIST:FREQuency?
Return data1.000000E+03(Format is in <NR3)
<Command>
:LIST:VOLTage <voltage NR3/disc >
:LIST:VOLTage?
FunctionIt sets or queries the measurement voltage of list mode.
Description
Set parameterWhen RO(output impedance) is setting 100Ω, the setup range of AC voltage is
{0.01~2|MAXimum|MINimum}. When RO(output impedance) is setting 25Ω, the setup range
of AC voltage is {0.01~1|MAXimum|MINimum}. The range of DC voltage is
{0.01~1|MAXimum|MINimum}.
Set syntax
:LIST:VOLTage 1
:LIST:VOLTage 1M
:LIST:VOLTage 1MV
:LIST:VOLTage 1E-3
:LIST:VOLTage MAXimum
:LIST:VOLTage MINimum
Query syntax
:LIST:VOLTage?
Return data1.000000E+00 (Format is in <NR3>)
When the Level setup is current mode, the data returns 9.9E37.