Water Level Sensor
Location / Function
During the Fill Mode...
5 VDC is supplied to the water level sensor. When the water level in the
chamber reaches the sensor disk, a circuit is completed, and current ows
back to the PC board.
When the 5 VDC from the water level sensor is detected, the PC board
stops the current ow to the ll valve.
Approximate quantities of water to reach water level sensor:
M9 units ..............650 ml
M11 units ............750 ml
During the Heat-Up / Sterilization / Vent / Dry Modes...
e water level sensor is not monitored.
Water Level Sensor
(rear view)
* Service Tip
Residue build-up on the sensor disk can cause malfunctions
during the Fill mode. Clean the sensor disk with Speed-Clean
and an abrasive pad, then dry disk thoroughly.